SPECIFICATIONS |
|
INPUT |
Voltage Range |
90-264VAC |
Frequency |
47-63Hz |
Power Factor |
EN61000-3-2 Amendment 14 Compliant |
Inrush Current |
35A Max (cold start) |
Leakage Current |
100μA max @ 264VAC |
Current @ 115VAC |
2.5A max |
Fusing |
3.15A / 250VAC |
|
OUTPUT |
Power | 65W Convection |
85W with 15 cfm Airflow |
Hold-up Time | 16mSec |
Efficiency |
75% Typical |
Adjustment Range (V1 Only) | ±5% |
Ripple / Noise, max. |
1% pk-pk max. |
Line Regulation |
±0.2% max. |
Load Regulation @ 60% ±40% Full Load |
V1: ±3% max.
V2-V4: ±5% max. |
Cross Regulation @ 60% ± 40% Full Load |
V1: Change in V2 - V4: ±0.5%
V2 - V4: Change in V1 @75 ±25% F/L: ±5% max. |
Overvoltage Protection (V1 Only) |
>125% (Latch Off) |
Power Limit |
>120% (Auto-Recovery) |
Overshoot (all outputs) |
10% max. |
Response Time |
500μSec (25-75% step load) |
Switching Frequency |
60KHz (typical) |
|
ENVIRONMENTAL |
Operating Temp. Range | -20°C to +50°C (Full Load)
Consult factory for -40°C Guaranteed Start-Up
and Industrial Temperature Range options |
Output Current Derating | 2.5%/°C, 50°C to 70°C |
Storage Temp. Range |
-40°C to + 85°C |
Humidity |
5% to 95%, Non-Condensing |
MTBF, Demonstrated |
>500,000 Hours |
Cooling |
15 cfm Airflow for Full Power |
Immunity | EN61000-4-2; -3; -4; -5; -6; -8; -11 |
Altitude | 10,000 feet |
|
PHYSICAL SPECIFICATIONS |
Case Dimensions | 5.00 x 3.00 x 1.20” / 127 x 76.2 x 30.48mm |
Weight | 0.75 lbs. (0.34 kg.) |
Vibration from 10 - 55Hz | 1.0G Peak
.(3 orthogonal axes @ 1 octave/min, 5 minute dwell @ 4 major resonances) |
|
COMPLIANCE |
EMC | FCC Class A & VDE Class A, CISPR 11; EN 55011 Class A |
Safety | UL60601-1 3rd Ed., EN60601-1 3rd Ed., CB REPORT (IEC 60601-1), CE MARK (LVD) |
“IF WHAT YOU SEE IS WHAT YOU DON’T WANT, IT CAN EASILY BE CHANGED.”
The MEDIMOD® family of switched mode medical power supplies has been designed with two precepts; (1) the laws of physics are immutable, and (2) the satisfaction of customer requirements and needs is paramount.
A host of modifications, only some of which are listed below, can and will be performed on products for customer programs requiring as few as 250 units per year. These “mods” are available at nominal premium (if any), normally without non-recurring engineering costs (although a one time documentation fee may be incurred), and usually with all safety agency approvals in place. This minimizes both product development cost and new product time to market. Effectively, MEDIMODs® medical switchers allow small program requirements the luxury of costly custom power supply designs.
TYPICAL MODIFICATIONS
- Unique Output Combinations from 1.5 to >48 volts
- Isolated Outputs
- Low Output Ripple and Noise
- Extended Temperature Operating Range
- -40°C Start-Up
- Zero Load Operation
CALL: +1-954-905-1070
EMAIL: the.power.solution@unipowerco.com
Looking for the DC input version of this product? Click HERE.
Looking for the ITE Approved version of this product? Click HERE.